Instruction Manual for TESA CLINOBEVEL 3
35
m
m
c
eff
g
g
sin
arcsin
whereas
g
c
Gravity at the location of calibration
α
m
Angle displayed at measuring site
g
m
Gravity at the measurement location
α
eff
Effective angle
In order to switch the correction of the local gravity on respectively off, proceed as follows:
Using the key
select the menu point [Options] and confirm your selection with
. Select now [Gravity] and confirm with
.
Switch the correction of the gravitation on using the keys
and confirm with the key
.
Now you can enter the value of the local gravity. The
value is adjusted with the keys
.
With the key
the standard value
9.807 m/s
2
will be recalled.
Confirm your entry with the key
.
The measuring instrument returns to the measuring
mode.
8.10 Firmware
Version
With this option information about the installed firmware and the configuration can be
displayed.
Using the key
select the menu point [Options] and confirm your selection with
. Select now [Version] and confirm with
.
The following information will be displayed:
Serial number of the instrument
Number of the firmware
Firmware release date
Instrument type (wireless / cable)
Serial number of the integrated sensor
Measuring range
Quick Calibration (ON/OFF)
After 10 seconds or with the key
this display will be left.
The measuring instrument returns to the measuring mode.
8.11 Reset quick
calibration
The data of the quick calibration will be deleted and replaced by the factory default values.
Using the key
select the menu point [Options] and confirm your selection with
. Select now [Reset Quick Calibration] and confirm with
.
9.80700
Local Gravity
m/s
2