Instruction Manual for TESA CLINOBEVEL 3
36
In order to prevent a resetting by error the question
"Are you sure?" will appear. If you really want to
delete the data of the quick calibration, press now the
key
. After 10 seconds or with the key
the instrument will return to the
measuring mode.
8.12 Reset to factory
settings
A complete factory reset will reset the instrument to the state as it has been configured at the
factory. All personal settings get lost.
The TESA CLINOBEVEL 3 will be set to the following standard configuration:
Standard values:
Measuring mode:
absolut
measuring unit:
DEC xx°xx'xx"
relative base:
1000 mm
absolute Zero point (ZERO-OFFSET):
0
relative Zero point (REL ZERO-OFFSET):
0
Filter No. 3
Display vial
Limits
OFF;
Upper Limit
0
Lower Limit
0
Scale
maximum range
Pairing
not paired
PIN code
OFF; Code = 00000
Display settings
Colour blue,
Saturation 100%,
Saturation
Power saving mode 50%
Logarithmic scaling
OFF
Function keys
all keys enabled;
Switching functions on or off
all keys enabled;
Hide disabled functions OFF
Wireless connection
ON, if available
Gravitation
OFF; Value = 9.807
kg·m/s
2
Using the key
select the menu point [Options] and confirm your selection with
. Select now [Factory Reset] and confirm with
.
In order to prevent a resetting by error the question
"Are you sure?" will appear. If you really want to
delete all personal settings, press now the key
. After 10 seconds or with the key
the instrument will return to the
measuring mode.
8.13 Function check
At the start of the instrument a system test will be performed, checking the most important
functions. In addition to this test the function of the keys and of the LEDs can be checked.
Using the key
select the menu point [Options] and confirm your selection with
. Select now [Self Test] and confirm with
.
Reset Quick Calibration
Are you sure?
Factory Reset
Are you sure?