EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
88
Changing the user code
1
Cabinet in work mode.
See
§5.4 Starting up the cabinet
on page 26.
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off
2
Change the user code
•
Push button
(Menu) and
(C).
•
Push 4x button
(=>).
•
Push button
.
•
Enter actual user code.
•
Push button
.
•
Enter the new valid user code.
•
Push button
.
User code:
0000 (Enter code)
A = Enter
>>
+
-
Esc
Input user code:
New code: 0000
Old code: 0001 A = Enter
>>
+
-
Esc
The pre installed valid user code is 0001. In case of emergency there is a special code to break the unknown
or forgotten user code. In that case enter code 2882 and the original pre installed code, 0001, is valid again.