EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
41
Status information cabinet shows (see example).
•
Middle numbers are the actual air speeds (yellow);
•
Left and right numbers are the set up alarm borders
(red).
4
Information fan shows
•
Push button
(=>).
The seven-digit number in the upper right corner is only
meant for service aims.
•
Push button
(Esc) for starting menu.
Information fan setting (0-1000) 1110011
Off: 0 Power save: 350
Work mode: 520 Actual: 537
=>
<=
Esc
*
1
The average inflow velocity at the top of the front aperture is measured and can be read in the display. The minimal required value
of the average inflow velocity measured at the top of the front aperture is given in chapter 9.2 on page 61. Measuring these most
critical measured values directly under the window ensures that the overall average inflow velocity is according to the guidelines of the
EN12469:2000 (Annex H table H1 Mean inflow velocity to achieve operator protection > 0,4 m/s).
Information about the installed options
(The display only shows when the options are installed)
1
Cabinet in work mode.
See
§5.4 Starting up the cabinet
on page 26.
•
Push button
(Menu).
•
Push button
(B).
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off
A => Hour counter information
B => Status cabinet
C => Settings
A
B
C
Esc
2
Information about the installed
options is given (example):
•
Push button
(=>).
Information status cabinet
A = Enter
=>
<=
Esc
Installed Options
1 Socket present, UV inst.
Gas present, 2 flowsens., Biohazard:Yes
=>
<=
Option
Esc
From here there are 3 possibilities (examples):
-
3a. Screen I
-
3b. Screen II
-
3c. Screen III