EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
57
Supplies needed
:
Usermanual;
Plastic bag (minimum measurement 75x50 [cm] with closure);
Two pairs of gloves;
Tape;
Working surface hook (auxiliary hook);
Possible disposing requirements;
New pre-filter (measurements of the pre-filter are necessary when ordering a new one, the
measurements are printed on the type plate (see
§4.1 Type plate on page 14
and
Appendix
VIII: Stickers on page 81
).
Figure 28 Supplies needed
Procedure
:
Switch on the installation;
Make sure that the window is in the working position and that there are no alarms;
Put on the gloves and keep an extra pair ready for use on the aerofoil;
Fold the edge of the plastic bag
±
10 cm;
Figure 29 Folding the edge of the bag
±
10 cm
Lift up the worktop with your left hand using the working surface hook and put it against the
back partition with your right hand;