EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
26
All equipment and, if possible, all materials required for working with a safety cabinet need to be
placed into the installation before starting up the cabinet. Equipment or objects with a height up to
350 mm can be placed in the working space by sliding the window upwards. In case of larger
objects the window can be hinged. Be careful: keep the safety regulations in consideration.
5.4
Starting up the cabinet
1
The lighting of the display will
light up.
The alarm indicator light is red
.
•
Connect the power cable.
Cabinet switched off
–
not safe
Monday 11:06:33
MENU
Socket=Off
2
Use both handgrips when hinging
or sliding the window.
•
Put the window in the working position (see marks
on the installation
).
3
The fan starts.
Cabinet is in start up mode.
The alarm indicator light is red
.
There is an acoustic alarm.
•
Push button
.
•
To acknowledge this alarm, push button
.
Starting to work mode
–
not safe
Monday 11:06:33
MENU
Socket=Off
4
If the cabinet is secured with a
user code, this code has to be
entered first.
The default code is 0001.
The user code can be changed
by the user, see
Appendix X:
User code on page 87.
•
Enter actual user code.
•
Confirm the code with button
.
User code:
0000 (Enter code)
A = Enter
>>
+
-
Esc
5
The light will switch on.
•
Push button
.
6
When the fan has reached the
right airspeed, the alarm
indicator will disappear.
The following display will be
shown.
However, the cabinet must operate for minimally 15
minutes more, before the cabinet is ready to work with.
Cabinet in work mode
–
safe
Monday 11:06:33
MENU
Socket=Off