EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
65
10
Trouble shooting
Fault Cause
Remedy
Socket is not activated
Activate socket on display push on
No power on socket of the cabinet
Fuse is broken
Replace the fuse
UV equipment is not installed
x
UV button is not visible on display
Option UV is not installed
UV option has to be programmed please
contact your service organization
UV is not working
Window not down
Window completely downwards
Gas equipment is not installed
x
Gas button is not visible on display
Option gas is not installed
UV option has to be programmed please
contact your service organization
Fan in off position
Gas is not working
High flow alarm/ Alarm in safe/night –
mode.
Fan in work mode
Contacts of the FL are not locking in the
power connections
Turn the FL-light slightly
FL-light is not working
UV light is on
Switch of the UV light
Cabinet is locked
User code must be entered
Enter the user code (or de-activate the user
code see
§5.14 Change the user code on
page 45).