EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
16
4.9
Drip tray/pre-filter (only EF/B and as an option for the EF)
The pre-filter with drip tray (EF:
Figure 3
and
EF/B:
Figure 5,
Pos. 9) is assembled under the
worktop. The installation is equipped with this to protect the interior of the installation against
pollution by coarse materials and to collect spilled fluids. Furthermore the pre-filter will enhance
the life span of the HEPA filter significantly.
4.10
Downflow HEPA filter
The downflow HEPA filter (
Figure 3, Figure 4, Figure 5
and
Figure 6,
Pos. 10) has a very high
efficiency: Class H14 in accordance with EN1822. After assembling, each filter will be tested
separately according to EN12469:2000, Annex D. Through the service panel the downflow filter
can be reached (Pos. 15).
4.11
Downflow protection
The downflow filter has a build-in protection grid (
Figure 3, Figure 4, Figure 5
and
Figure 6,
Pos.
11).
4.12
Exhaust HEPA filter
The exhaust HEPA filter (
Figure 3, Figure 4, Figure 5
and
Figure 6,
Pos. 12) has a very high
efficiency: Class H14 in accordance with EN1822. After assembling, each filter will be tested
separately according to EN 12469:2000, Annex D. Through the topside of the cabinet the exhaust
filter can be reached.
4.13
Support frame (optional)
For a stable positioning of the cabinet, we strongly recommend the use of the especially designed
support frame from the supplier (
Figure 3, Figure 4, Figure 5
and
Figure 6,
Pos. 13).
This support frame includes an in steps adjustable feet support for a good ergonomic work
position (
Figure
3
, Figure 4, Figure 5
and
Figure 6,
Pos. 14). As well, the feet support will increase
the stability of the cabinet. In addition, it is recommended to use an in height adjustable, well
disinfectable chair to optimize the work position.
4.14
Feet support (assembled on 4.13)
The support frame includes an in steps adjustable feet support for a good ergonomic work position
(
Figure 3, Figure 4, Figure 5
and
Figure 6,
Pos. 14). As well, the feet support will increase the
stability of the cabinet. In addition, it is recommended to use an in height adjustable, well
disinfectable chair to optimize the work position.
4.15
Front panel
The service panel (
Figure 3, Figure 4, Figure 5
and
Figure 6,
Pos. 15) can be dismantled to get
access to:
•
the downflow filter;