EuroFlow EF and EF/B
Biological Safety Cabinet
UM-EFAB-EN10.doc
34
-
2c. Disable
Timer
2a
The timer sets a period until the
UV turns off.
•
Push button
(Timer), to set up the timer.
•
Push button
(Change), to change the time.
•
The value can be changed by pushing button
(>>),
(+) or
(-).
•
Acknowledge the setting with button
.
•
Push 2x button
(Esc) for starting menu.
Clock (7 days clock
)
2b
With the clock function, for each
day of the week, you can set a
variable period for the UV to be
switched on. Keep in mind, you
have to set the starting hour and
starting minute for each day
separately! For detailed
information regarding
programming this menu, see
§5.11 Set up 7 days clock UV on
page 33
.
•
Push button
(Clock), to set up the 7 days clock.
•
Push button
(Change), to change starting time en
ending time.
•
The value can be changed by pushing button
(>>),
(+) or
(-).
•
Acknowledge the setting with button
.
•
Push 2x button
(Esc) for starting menu.
See also the
Example on page 34
.
Disable
2c
By pushing on “Disable” the UV
can only be turned on and off
manually.
•
Push button
(Disable), to set the UV manually.
•
Push 2x button
(Esc) for starting menu.
Please select not only a starting time, but also an ending time for every programmed day.
Otherwise the UV will stay on until 00:00 h and only then turn off. See also the
Example on page
34.
Example
Programming the 7 days clock UV will be given as an example.
Switch on the UV Tuesday evening 22.00 and switch off the UV Wednesday morning 06.00.
(See
§5.11 Hour counter information and reset (FL, UV,
Timer)
point 5a on page 32,
for deleting
all Hour counter programmed information).
1
Cabinet in work mode.
See
§5.4 Starting up the cabinet
on page 26.
•
Push button
(Menu).
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off