Performance Verification
2- 40
WFM700 Series Waveform Monitors Service Manual
10.
Set
Hi Pass Filter
to
1 kHz
.
11.
Change RF generator settings as follows:
LFGEN2 Frequency
1 kHz
12.
Repeat steps 7 through 9.
13.
Set
Hi Pass Filter
to
10 kHz
.
14.
Change RF generator settings as follows:
LFGEN2 Frequency
10 kHz
15.
Repeat steps 7 through 9.
16.
Change RF generator settings as follows:
LFGEN2 Frequency
100 kHz
17.
Set
Hi Pass Filter
to
1 kHz
.
18.
Adjust the RF generator deviation for a Jitter reading of 4.0 UI.
19.
Set
Hi Pass Filter
to
100 kHz
.
20.
Repeat steps 7 through 9.
Summary of Contents for WFM700 Series
Page 4: ......
Page 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Page 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Page 19: ...Theory of Operation...
Page 20: ......
Page 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Page 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Page 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Page 107: ...Adjustment Procedures...
Page 108: ......
Page 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Page 133: ...Maintenance...
Page 134: ......
Page 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Page 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Page 177: ...Replaceable Parts...
Page 178: ......
Page 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...