Adjustments
3- 14
WFM700 Series Waveform Monitors Service Manual
Adjust signal to cursors,
disregard peaks and dips
Figure 3- 10: Eye vertical scale adjustment screen
16.
Touch the
Save and Exit
soft key to save the new calibration constant.
17.
Touch the
CONFIG
button twice to exit the configuration menu.