Adjustments
3- 10
WFM700 Series Waveform Monitors Service Manual
13.
Move the cable from the
PIX B/Pb
to the
PIX R/Pr
connector.
14.
Adjust R843 for a display of 700 mV on the test oscilloscope. Measure the
peak-to-peak amplitude of the waveform displayed on the oscilloscope. See
Figure 3--7.
Black level
First color bar
Ground reference
Figure 3- 6: HD PIX G/Y output adjustment screen
Black level
Pk - Pk level
Figure 3- 7: HD PIX B/Pb and PIX R/Pr output adjustment screen
Summary of Contents for WFM700 Series
Page 4: ......
Page 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Page 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Page 19: ...Theory of Operation...
Page 20: ......
Page 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Page 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Page 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Page 107: ...Adjustment Procedures...
Page 108: ......
Page 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Page 133: ...Maintenance...
Page 134: ......
Page 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Page 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Page 177: ...Replaceable Parts...
Page 178: ......
Page 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...