Performance Verification
2- 4
WFM700 Series Waveform Monitors Service Manual
Performance Verification Conditions
NOTE
. Menu selections in the WFM700 Waveform Monitor differ between
firmware versions. The
Performance Verification
procedures assume you have
the latest firmware version installed in your instrument.
The following conditions must be met prior to doing these procedures:
1.
The WFM700 Waveform Monitor must have been operating continuously
for twenty (20) minutes in an environment that meets the operating range
specifications for temperature and humidity.
2.
Connect the WFM700 Waveform Monitor and the test equipment to the
same AC power circuit or to a common power strip if you are unsure of the
AC power circuit distribution.
Self Tests
This procedure uses internal routines to verify that the waveform monitor
functions and passes its internal self tests. No test equipment or hookups are
required.
1.
Press the
PRESET
button.
2.
Touch the
--more--
soft key, then touch the
Restore Factory
soft key.
NOTE
. Restoring factory settings overrides the current instrument state but does
not alter calibration or presets.
3.
Press the
CONFIG
front panel button.
4.
Touch
System
and then
Diagnostics
.
5.
Touch
Diagnostics
, then
View Diagnostic Log
.
6.
Touch
**Erase** NVRAM Log
to clear the log file.
7.
Touch
Back to Diags Menu
.
8.
Touch
Run Full Powerup Diagnostics
.
9.
Touch
Back to Diags Menu
.
10.
Touch
View Diagnostics Log
.
11.
CHECK
the diagnostics log pages to verify that there are no red FAIL
messages.
Summary of Contents for WFM700 Series
Page 4: ......
Page 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Page 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Page 19: ...Theory of Operation...
Page 20: ......
Page 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Page 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Page 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Page 107: ...Adjustment Procedures...
Page 108: ......
Page 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Page 133: ...Maintenance...
Page 134: ......
Page 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Page 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Page 177: ...Replaceable Parts...
Page 178: ......
Page 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...