Adjustments
3- 22
WFM700 Series Waveform Monitors Service Manual
Performing this procedure loads new calibration constants into the instrument.
This calibration routine is only available for the WFM700M (WFM700 with a
WFM7M module).
The following test items are required to perform the entire adjustment.
Adjustment requirements
Recommended test equipment and test signal
Calibrated sine wave generator,
741.7582 MHz, 800 mV into
75
8
, FM modulated to generate
2 UI jitter amplitude
RF generator adjusted to 800 mV output into 75
8
at the
requested frequency.
With the recommended RF generator, you can modulate the
signal with the following settings to obtain 2 UI of jitter.
Frequency
741.7582 MHz
Output Level
0 dB
Frequency Modulation
OFF
LFGEN2 Frequency
10 Hz
LFGEN2 Waveshape
SINE
Phase Modulation Source
LF2
Phase Deviation
6.28 RADIANS
RF Output
ON
Phase Modulation
ON
75
8
coaxial cable
36 inches long
NOTE
. If this test is performed out of sequence or the sequence of tests has been
interrupted (such as power off or input from another module was selected),
perform the steps in this note to prepare the instrument to continue testing.
a.
Connect the SD video generator to
INPUT B
using a 75
8
cable.
b.
Set the SD video generator for a 525-format 100% color bar signal.
c.
Press the
INPUT
button and select
Input B
.
d.
Press the
EYE
button.
e.
Press the
INPUT
button and select
Input A
.
1.
Install the 50
8
N-to-BNC adapter on the RF generator output.
2.
Connect the RF generator output to
INPUT A
using a 75
8
cable. Imped-
ance matching is not important for this connection.
3.
Touch the
Jitter HPF Adjust
soft key and then the
Start Jitter HPF
Adjustment
soft key.
4.
Adjust the displayed Measured Jitter amplitude using the General Purpose
Knob for a reading of 2.83 UI. See Figure 3--14.
5.
Touch the
Save and Exit
soft key to save the new calibration constant.
Jitter HPF Adjust
Summary of Contents for WFM700 Series
Page 4: ......
Page 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Page 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Page 19: ...Theory of Operation...
Page 20: ......
Page 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Page 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Page 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Page 107: ...Adjustment Procedures...
Page 108: ......
Page 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Page 133: ...Maintenance...
Page 134: ......
Page 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Page 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Page 177: ...Replaceable Parts...
Page 178: ......
Page 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...