Adjustments
WFM700 Series Waveform Monitors Service Manual
3- 19
This procedure sets the calibration constant for the eye diagram clock recovery
bandwidth. This calibration routine is only available for the WFM700M
(WFM700 with a WFM7M module).
Following is a list of the signal requirements to perform the adjustment.
Adjacent to the requirement is the recommended equipment and test signals.
Adjustment requirements
Recommended test equipment and test signal
Calibrated sine wave generator,
741.7582 MHz, 800 mV into
75
8
, with 10 Hz jitter
modulation
With the recommended RF generator, you can generate the
required signal with the following settings to obtain 10 Hz
jitter modulation.
Frequency
741.7582 MHz
Output Level
0 dB
LFGEN2 Frequency
10 Hz
LFGEN2 Waveshape
SINE
Phase Modulation Source
LFGEN2
Phase Deviation
2.22 RADIANS
RF Output
ON
Phase Modulation
ON
75
8
coaxial cable
36 inches long
NOTE
. If this test is performed out of sequence or the sequence of tests has been
interrupted (such as power off or input from another module was selected),
perform the steps in this note to prepare the instrument to continue testing.
a.
Connect the SD video generator to
INPUT B
using a 75
8
cable.
b.
Set the SD video generator for a 525-format 100% color bar signal.
c.
Press the
INPUT
button and select
Input B
.
d.
Press the
EYE
button.
e.
Press the
INPUT
button and select
Input A
.
1.
Install the 50
8
N-to-BNC adapter on the RF generator output.
2.
Connect the RF generator output to
INPUT A
using a 75
8
cable. Imped-
ance matching is not important for this connection.
3.
Touch the
Eye Loop BW Adjust (10 Hz)
soft key and then the
Start Eye
Loop BW Adjustment (10 Hz)
soft key.
4.
Adjust the RF generator output level for a displayed waveform amplitude of
approximately 800 mV.
5.
Adjust the displayed modulation using the General Purpose Knob for exactly
1 UI as measured on the display. See Figure 3--13.
6.
Touch the
Save and Exit
soft key to save the new calibration constant.
Eye Loop BW Adjust
(10 Hz)
Summary of Contents for WFM700 Series
Page 4: ......
Page 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Page 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Page 19: ...Theory of Operation...
Page 20: ......
Page 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Page 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Page 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Page 107: ...Adjustment Procedures...
Page 108: ......
Page 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Page 133: ...Maintenance...
Page 134: ......
Page 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Page 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Page 177: ...Replaceable Parts...
Page 178: ......
Page 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...