Adjustments
WFM700 Series Waveform Monitors Service Manual
3- 11
Instrument Driven Procedures
The following procedures use the WFM700 internal calibration routines to load
calibration constants into non-volatile memory. The routines automatically make
the necessary instrument settings. The user must step through the procedure and
provide the signal sources to complete the routine. See Figure 3--8 for the menu
structure of the calibration routines.
Internal access to the instrument is not required for these procedures.
Calibration
Touchpanel Calibration
Eye Gain Adjust
SD Jitter Gain Adjust
HD Jitter Gain Adjust
Eye Loop BW Adjust (10 Hz)
Eye Loop BW Adjust (100 Hz)
Eye Loop BW Adjust (1 kHz)
Jitter HPF Adjust
Eye Sig BW Test
These menu select ions appear only
on the WFM700M (WFM700 with a
WFM7M video module)
Figure 3- 8: Calibration routines menu
The calibration routines are found in the configuration menus. Read these
general instructions to familiarize yourself with the location of the menus and
how to start and complete each routine.
a.
Perform the procedures in the order presented.
b.
Navigate to the calibration routines by pressing the
CONFIG
button and
then the
Calibration
soft key.
c.
The on-screen provide brief instructions for signal requirements and
setup. Refer to the procedural steps in this section for a description of the
test, signal requirements, and operational hints for using the recom-
mended test equipment.
d.
Each routine requires that you either save your new calibration constants,
revert to the previously saved value, or set to default values.
General Instructions
Summary of Contents for WFM700 Series
Page 4: ......
Page 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Page 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Page 19: ...Theory of Operation...
Page 20: ......
Page 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Page 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Page 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Page 107: ...Adjustment Procedures...
Page 108: ......
Page 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Page 133: ...Maintenance...
Page 134: ......
Page 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Page 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Page 177: ...Replaceable Parts...
Page 178: ......
Page 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...