Calibration
WCA230A & WCA280A User Manual
1- 25
Calibrating DC Offset
The DC offset calibration cancels DC offset that appears at 0 Hz in the baseband
(DC to 20 MHz). When you change the amplitude setting and the DC offset is
too obvious, run the DC offset calibration.
When DC offset appears at 0 Hz in
the baseband, run the calibration.
Figure 1- 20: DC offset
To run the DC offset calibration, do the following:
1.
Press the
CAL
key on the front panel.
2.
Press the
Calibrate DC Offset
side key.
The calibration runs. It takes several seconds to complete the process.
Summary of Contents for WCA200A
Page 6: ......
Page 8: ......
Page 29: ...Getting Started...
Page 30: ......
Page 57: ...Operating Basics...
Page 58: ......
Page 86: ...Functional Overview 2 28 WCA230A WCA280A User Manual...
Page 125: ...Reference...
Page 126: ......
Page 208: ...Modulation Analysis Demod Mode 3 82 WCA230A WCA280A User Manual...
Page 272: ...Setting Timing Parameters 3 146 WCA230A WCA280A User Manual...
Page 346: ...Display Line 3 220 WCA230A WCA280A User Manual...
Page 358: ...Marker Operation and Peak Search 3 232 WCA230A WCA280A User Manual...
Page 388: ...File Format 3 262 WCA230A WCA280A User Manual...
Page 392: ...Screen Copy 3 266 WCA230A WCA280A User Manual...
Page 393: ...Appendices...
Page 394: ......
Page 430: ...Appendix C Digital Demodulation Symbol Mapping C 4 WCA230A WCA280A User Manual...
Page 437: ...Glossary...
Page 438: ......
Page 446: ...Glossary Glossary 8 WCA230A WCA280A User Manual...
Page 447: ...Index...
Page 448: ......