![Tektronix WCA200A User Manual Download Page 237](http://html1.mh-extra.com/html/tektronix/wca200a/wca200a_user-manual_1078248237.webp)
Time Analysis (Time Mode)
WCA230A & WCA280A User Manual
3- 111
Measures the phase noise in real time. The measurement results can be displayed
in three dimensions like spectrogram, called “Noisogram”: the color axis shows
the C/N (dBc/Hz), the horizontal axis shows the offset frequency (Hz) and the
vertical axis shows the time (frame number). Jitter equivalent to the RMS value
of the phase noise and the settling time of the phase noise and jitter are also
calculated from the phase noise data.
Meas Setup Menu.
The following items are provided in the Meas Setup menu for
the Real-Time Phase Noise measurement.
Analyze.
Performs measurement for acquisition data in the analysis range.
NOTE
. When you change settings in the Meas Setup menu, press the
Analyze
side key to perform the measurement for the modified settings.
The next items are the same as for the phase noise measurement.
Refer to page 3--105.
Carrier Threshold Level
Carrier Bandwidth
C/N Sideband
Rj Start Offset Frequency
Rj Stop Offset Frequency
Max Pj Threshold
The following items are added for the real-time measurement (see Figure 3--77).
They depend on the subview content (refer to
Selecting the Subview Content
on
page 3--113).
Rj Settling Threshold.
Enabled when Random jitter versus Time is displayed
in the subview. Sets the threshold value for obtaining the random jitter settling
time.
Setting range: 0 to 1 s (default: 0 s).
C/N Offset Frequency.
Enabled when C/N versus Time is displayed in the
subview. Sets the value of the frequency displaying the C/N versus Time in the
subview as the offset from carrier frequency.
Setting range: The upper limit is Span/2. The lower limit is determined by the
span and FFT Points for C/N (refer to
Setting the Number of FFT Points
on
page 3--112) settings.
C/N Settling Threshold.
Enabled when C/N versus Time is displayed in the
subview. Sets the threshold value for obtaining the phase noise settling time.
Setting range: --200 to 0 dBc/Hz (default: 0 dBc/Hz).
Real-Time Phase Noise
Measurement
Summary of Contents for WCA200A
Page 6: ......
Page 8: ......
Page 29: ...Getting Started...
Page 30: ......
Page 57: ...Operating Basics...
Page 58: ......
Page 86: ...Functional Overview 2 28 WCA230A WCA280A User Manual...
Page 125: ...Reference...
Page 126: ......
Page 208: ...Modulation Analysis Demod Mode 3 82 WCA230A WCA280A User Manual...
Page 272: ...Setting Timing Parameters 3 146 WCA230A WCA280A User Manual...
Page 346: ...Display Line 3 220 WCA230A WCA280A User Manual...
Page 358: ...Marker Operation and Peak Search 3 232 WCA230A WCA280A User Manual...
Page 388: ...File Format 3 262 WCA230A WCA280A User Manual...
Page 392: ...Screen Copy 3 266 WCA230A WCA280A User Manual...
Page 393: ...Appendices...
Page 394: ......
Page 430: ...Appendix C Digital Demodulation Symbol Mapping C 4 WCA230A WCA280A User Manual...
Page 437: ...Glossary...
Page 438: ......
Page 446: ...Glossary Glossary 8 WCA230A WCA280A User Manual...
Page 447: ...Index...
Page 448: ......