![Tektronix WCA200A User Manual Download Page 255](http://html1.mh-extra.com/html/tektronix/wca200a/wca200a_user-manual_1078248255.webp)
Setting Amplitude
WCA230A & WCA280A User Manual
3- 129
RF Atten/Mixer.
Input signals are attenuated through the attenuator and
converted to IF signals through the mixer in the down-converter (refer to
Architecture
on page 1--6). The attenuation level and the mixer level are normally
set automatically. Select
RF Att
or
Mixer
to manually set either parameter, if
necessary. RF Atten/Mixer is unavailable in the baseband (DC to 20 MHz).
H
Auto.
Sets the mixer level and the RF attenuation level automatically.
H
RF Att.
Sets the RF attenuation level with
RF Att
described below.
H
Mixer.
Sets the mixer level using the
Mixer Level
setting.
NOTE
. RF Atten/Mixer is set to Auto by default. In this setting, the mixer level is
fixed to --25 dBm.
RF Att.
Changes the RF attenuation level when
RF Att
is selected in
RF Atten/Mixer
above.
Table 3- 19: RF attenuation level setting
Frequency band
Attenuation level (dBm)
RF
(WCA230A)
RF1
(WCA280A)
0 to 50 (2 dB step)
RF2, 3 (WCA280A)
0, 10, 20, 30, 40, or 50
Reducing the attenuation may increase the signal level compared to the noise
floor, because the noise is generated after the RF attenuator.
Mixer Level.
Selects the input level of the primary mixer when
Mixer
is
selected in
RF Atten/Mixer
.
Table 3- 20: Mixer level setting
Frequency band
Mixer level (dBm)
1
RF
(WCA230A)
RF1
(WCA280A)
--5, --10, --15, --20, or --25
RF2, 3 (WCA280A)
--5, --15, or --25
1
- 25 dBm fixed when RF Atten/Mixer is set to Auto.
Select the level according to measurement type. The default value is --25 dBm.
Use the default value in most cases. This level can be increased up to --5 dBm
when a high dynamic range is required for measurements such as the ACPR
(Adjacent Channel Leakage Power Ratio) measurement.
Summary of Contents for WCA200A
Page 6: ......
Page 8: ......
Page 29: ...Getting Started...
Page 30: ......
Page 57: ...Operating Basics...
Page 58: ......
Page 86: ...Functional Overview 2 28 WCA230A WCA280A User Manual...
Page 125: ...Reference...
Page 126: ......
Page 208: ...Modulation Analysis Demod Mode 3 82 WCA230A WCA280A User Manual...
Page 272: ...Setting Timing Parameters 3 146 WCA230A WCA280A User Manual...
Page 346: ...Display Line 3 220 WCA230A WCA280A User Manual...
Page 358: ...Marker Operation and Peak Search 3 232 WCA230A WCA280A User Manual...
Page 388: ...File Format 3 262 WCA230A WCA280A User Manual...
Page 392: ...Screen Copy 3 266 WCA230A WCA280A User Manual...
Page 393: ...Appendices...
Page 394: ......
Page 430: ...Appendix C Digital Demodulation Symbol Mapping C 4 WCA230A WCA280A User Manual...
Page 437: ...Glossary...
Page 438: ......
Page 446: ...Glossary Glossary 8 WCA230A WCA280A User Manual...
Page 447: ...Index...
Page 448: ......