Tutorial
2- 42
WCA230A & WCA280A User Manual
2.
Acquire data in the single mode using the following steps:
a.
Press the
TRIG
key on the front panel.
b.
Press the
Repeat...
side key and select
Single
.
c.
Press the
RUN/STOP
key to acquire data. One waveform is acquired
and displayed every time you press the key.
3.
Press the
Repeat...
side key again and select Continuous to return to the
continuous mode.
When acquisition starts, “READY” is displayed in the status indicator,
indicating that the analyzer is ready for trigger.
Summary of Contents for WCA200A
Page 6: ......
Page 8: ......
Page 29: ...Getting Started...
Page 30: ......
Page 57: ...Operating Basics...
Page 58: ......
Page 86: ...Functional Overview 2 28 WCA230A WCA280A User Manual...
Page 125: ...Reference...
Page 126: ......
Page 208: ...Modulation Analysis Demod Mode 3 82 WCA230A WCA280A User Manual...
Page 272: ...Setting Timing Parameters 3 146 WCA230A WCA280A User Manual...
Page 346: ...Display Line 3 220 WCA230A WCA280A User Manual...
Page 358: ...Marker Operation and Peak Search 3 232 WCA230A WCA280A User Manual...
Page 388: ...File Format 3 262 WCA230A WCA280A User Manual...
Page 392: ...Screen Copy 3 266 WCA230A WCA280A User Manual...
Page 393: ...Appendices...
Page 394: ......
Page 430: ...Appendix C Digital Demodulation Symbol Mapping C 4 WCA230A WCA280A User Manual...
Page 437: ...Glossary...
Page 438: ......
Page 446: ...Glossary Glossary 8 WCA230A WCA280A User Manual...
Page 447: ...Index...
Page 448: ......