Basic Installation
Figure 9: IP Properties dialog box
4.
Select
Use the following IP address
.
5.
Enter the IP address
169.254.0.22
and Subnet mask
255.255.0.0
for the
controller. No Default gateway or DNS server information is required.
6.
Click
OK
and close the dialog boxes.
Con
fi
gure the Benchtop Mainframe IP Address to a Static Address.
To complete the setup, set the IP address of the benchtop mainframe as a static
address. You can do this one of two ways: Use the preset addresses described in
step 7, or specify an address using the
Changing the TLA7016 Factory Network
Settings
procedure. (See page 32,
Changing the TLA7016 Factory Network
Settings
7.
On the TLA7016 Interface Module, press and hold the RESET button until
the IP address 169.254.0.19 shows in the TLA7016 front panel display. (See
Figure 5 on page 9.) The front-panel display cycles every 2-3 seconds to the
next preset address. (See Table 3 on page 10.)
8.
Start the TLA application by double-clicking the TLA Application icon on the
controller. The TLA Connection dialog box appears. (See Figure 10.)
TLA7000 Series Logic Analyzers Installation Manual
13