Appendix A: Mainframe Power Information
CAUTION.
By placing too many modules in a TLA7016 Benchtop Mainframe
you can overload the power distribution system and damage the mainframe
or modules. To avoid this, ensure that your module con
fi
guration in a single
mainframe does not exceed 1000 W at line voltages less than 120 V nominal at
50 Hz or 60 Hz or at line voltages of 115 V at 400 Hz. Use the following table to
calculate the total power for your mainframe. (See Table 8.)
The power consumption depends on the number and type of modules installed
in the mainframes. To determine the total power consumption, perform the
following steps:
1.
Use the following table to determine the power consumption for each module.
2.
Add the power for each module to determine the total power consumption.
Table 8: Power for instrument modules
Module type
Power (Watts)
One benchtop mainframe
1
70
TLA7AA1
45
TLA7AA2
65
TLA7AA3
85
TLA7AA4
105
TLA7AB2
65
TLA7AB4
105
TLA7NA1
45
TLA7NA2
58
TLA7NA3
71
TLA7NA4
82
TLA7Q2
51
TLA7Q4
75
TLA7P2
50
TLA7P4
74
TLA7N1
45
TLA7N2
58
TLA7N3
71
TLA7N4
82
TLA7L1
55
TLA7L2
73
TLA7L3
94
TLA7L4
109
TLA7000 Series Logic Analyzers Installation Manual
87
Summary of Contents for TLA7000 Series
Page 1: ...TLA7000 Series Logic Analyzers Installation Manual www tektronix com 071 1747 03 ...
Page 72: ...Basic Installation 54 TLA7000 Series Logic Analyzers Installation Manual ...
Page 88: ...Merging Modules 70 TLA7000 Series Logic Analyzers Installation Manual ...
Page 104: ...Restoring and Reinstalling Software 86 TLA7000 Series Logic Analyzers Installation Manual ...