Restoring and Reinstalling Software
1.
Note which of the modules that may require new
fi
rmware.
If any modules are missing from the System window, you can verify the status
of the modules by doing the following steps:
a.
Select System Properties in the System menu.
b.
Click the Modules tab.
c.
Note the information under the Messages column for any installed
modules. If any modules indicate invalid or similar messages, you must
update the
fi
rmware for those modules.
2.
Turn off the logic analyzer and disconnect the power cord.
3.
Remove the modules from the logic analyzer.
4.
Locate the
fl
ash programming pins on the rear of each module. (See Figure 59
on page 81.)
5.
Install a jumper on the
fl
ash programming pins (use one of the spare jumpers
that came with your logic analyzer).
6.
Reinstall the module(s) in the logic analyzer.
7.
Reconnect the power cord and then power on the logic analyzer.
8.
Wait for the TLA application to start.
NOTE.
Any modules with the
fl
ash programming jumper installed will not display
in the System window.
9.
Exit the TLA application (and pattern generator application if you have it
installed on your instrument).
80
TLA7000 Series Logic Analyzers Installation Manual
Summary of Contents for TLA7000 Series
Page 1: ...TLA7000 Series Logic Analyzers Installation Manual www tektronix com 071 1747 03 ...
Page 72: ...Basic Installation 54 TLA7000 Series Logic Analyzers Installation Manual ...
Page 88: ...Merging Modules 70 TLA7000 Series Logic Analyzers Installation Manual ...
Page 104: ...Restoring and Reinstalling Software 86 TLA7000 Series Logic Analyzers Installation Manual ...