Basic Installation
Installing Modules
TLA700 Modules are fully compatible with the TLA7000 Series mainframes. This
section describes how to install TLA700 modules in your TLA7000 mainframe.
CAUTION.
To avoid damaging the instrument, do not install or remove any
modules while the instrument is powered on. Always power down the instrument
before installing or removing modules.
Logical Address Switches
Every module in the instrument must have a unique logical address; no two
modules can have the same address. Use the two rotary switches on the rear panel
of the modules to select the logical address. (See Figure 38 on page 42.)
The factory default switch setting is FF (Dynamic Auto Con
fi
guration). With
dynamic logical address auto con
fi
guration, the instrument automatically sets
the address to an unused value.
Normally, you would not want to change the default settings. However, you can
set the address to a
fi
xed static logical address value. You can select any static
addresses between 01 and FE hexadecimal (1 to 254 decimal). If you set the
logical address switches to any other setting than FF, you must verify that no two
modules (or devices) share the same address.
NOTE.
Do not set any module to logical address 00. Logical address 00 is
reserved exclusively for the controller.
TLA7000 Series Logic Analyzers Installation Manual
41
Summary of Contents for TLA7000 Series
Page 1: ...TLA7000 Series Logic Analyzers Installation Manual www tektronix com 071 1747 03 ...
Page 72: ...Basic Installation 54 TLA7000 Series Logic Analyzers Installation Manual ...
Page 88: ...Merging Modules 70 TLA7000 Series Logic Analyzers Installation Manual ...
Page 104: ...Restoring and Reinstalling Software 86 TLA7000 Series Logic Analyzers Installation Manual ...