Basic Installation
You can also verify more detailed functionality by running the self-calibration and
extended diagnostics.
NOTE.
Allow the mainframe and modules to warm up for 30 minutes before
running the self-calibration.
To run self-calibration and diagnostics, perform the following:
1.
Disconnect any probes attached to the modules.
2.
Select the System menu and click Calibration and Diagnostics.
3.
Run the self-calibration, and then the extended diagnostics by selecting the
proper tab. Results of the tests display on the individual property page.
To run self-calibration and external diagnostics for a pattern generator, perform
the following:
1.
Select the Pattern Generator System menu and click Calibration and
Diagnostics.
2.
Run the self-calibration, and then the extended diagnostics by selecting the
proper tab. Results of the tests display on the individual property page.
NOTE.
The time required to run the self-calibration on the logic analyzer modules
depends on the number of acquisition channels. Modules with a large number of
channels may take several minutes to run the self-calibration.
Checking the P68xx and
P69xx Logic Analyzer
Probes (Optional)
To check the P68xx or P69xx logic analyzer probes, connect the probes to a signal
source, start an acquisition, and verify that the acquired data is displayed in either
the listing or waveform windows.
To run the performance veri
fi
cation, refer to the performance veri
fi
cation
procedure in the
TLA7Axx Logic Analyzer Module Service Manual.
Checking the P64xx Logic
Analyzer Probes (Optional)
Connect the P64xx logic analyzer probes to a signal source, start an acquisition,
and verify that the acquired data is displayed in either the listing or waveform
windows. You can also use the Activity Indicators in the logic analyzer Setup
window to view signal activity at the probe tips.
NOTE.
If you connect probes to any channels other than the A2 and A3 groups,
you must de
fi
ne the groups and channels in the Setup window before acquiring
data on other probe channels.
TLA7000 Series Logic Analyzers Installation Manual
51
Summary of Contents for TLA7000 Series
Page 1: ...TLA7000 Series Logic Analyzers Installation Manual www tektronix com 071 1747 03 ...
Page 72: ...Basic Installation 54 TLA7000 Series Logic Analyzers Installation Manual ...
Page 88: ...Merging Modules 70 TLA7000 Series Logic Analyzers Installation Manual ...
Page 104: ...Restoring and Reinstalling Software 86 TLA7000 Series Logic Analyzers Installation Manual ...