Specifications – TekRanger 2 (serial nos. B052000 and above)
1–62
TFS3031 TekRanger/TekRanger 2 Service Manual
Table 1–67: Vertical System Measurements (850 nm Multimode, Options 01 and 03)
Characteristic
Performance Requirement
Supplemental Information
Linearity
±0.02 dB/dB typical
3Ćminute average
From 12 dB above RMS floor to 1.5
dB below maximum signal level.
Measurement Repeatability
TwoĆPoint
Splice Loss
LSA Fiber Loss
±0.1 dB/dB typical
3Ćminute average
For events 10 dB above 98% noise
floor, 95% confidence.
Reflectance
±2 dB typical
NonĆclipped reflection.
Table 1–68: Single-Event Distance Measurement Repeatability
1
(850 nm Multimode, Options 01 and 03)
Characteristic
Performance Requirement
Supplemental
Information
Range from Front
Panel (dB)
0 to 10
> 10
Reflective
2
typical
2.5 m
5 m
Nonreflective
3
typical
2.5 m
15 m
IntelliTrace mode
1
Multiply by 0.001% of range for timebase error. These data are for a 0.5 dB event with 95% probability.
2
Reflectance greater than -45 dB, 0.5 dB loss.
3
Reflectance less than -60 dB, 0.5 dB loss.
Summary of Contents for TekRanger 2 TFS3031
Page 5: ......
Page 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 270: ......
Page 271: ......