![Tektronix TekRanger 2 TFS3031 Service Manual Download Page 178](http://html1.mh-extra.com/html/tektronix/tekranger-2-tfs3031/tekranger-2-tfs3031_service-manual_1078196178.webp)
Performance Check – TekRanger 2 (serial nos. B052000 and above)
TFS3031 TekRanger/TekRanger 2 Service Manual
4–17
Power/charger adapter
Fiber test fixture
4. Check that loss at the front-panel connection is minimized (0.75 dB or less).
a. Press the
Setup
softkey to display the Test Setup menu, and set the
range to 1 km and the pulse width to 5 m. Press
Exit
.
b. Press and hold the
START/STOP
button to start a real-time acquisition.
c.
If front-panel loss is greater than 0.75 dB, clean the fiber connector and
the laser output port.