
Performance Check – TekRanger (serial nos. below B052000)
TFS3031 TekRanger/TekRanger 2 Service Manual
3–27
11. Push the
softkey to again select the first cursor as the
active cursor.
12. Position the active cursor on the steepest part of the rising edge of the pulse
of the event in question.
13. The Cursor A–B reading at top screen is the attenuation dead zone.
14. Position the two cursors so they are 1.5 dB down from the peak of the
reflection on either side of the pulse.
15. The Cursor A–B reading at top screen is the event dead zone.
16. Repeat steps 4 through 15 with Pulsewidth setups of 5 m and 1 m.
17. Power off the TFS3031 and disconnect the multimode dead-zone/event-
detection test fixture.
Summary of Contents for TekRanger 2 TFS3031
Page 5: ......
Page 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 270: ......
Page 271: ......