Performance Check – TekRanger 2 (serial nos. B052000 and above)
TFS3031 TekRanger/TekRanger 2 Service Manual
4–41
Singlemode Reflectance-Accuracy Check (Options 04, 06, and 10)
1. Clean the connector on the TFS3031 and the TFS3031 singlemode reflec-
tance test fixture.
2. Connect the TFS3031 to the reflectance test fixture and turn on power to the
instrument.
3. Check that loss at the front-panel connection is minimized (0.75 dB or less).
a. Press the
Setup
softkey to display the Test Setup menu, and set the
range to 1 km and the pulse width to 5 m. Select 1550 wavelength, or
1310 if option 04. Press E
xit
.
b. Press and hold the
START/STOP
button to start a real-time acquisition.
c.
If front-panel loss is greater than 0.75 dB, clean the fiber connector and
the laser output port.
4. Set the reflectance test fixture for -40 dB at the wavelength of the laser being
tested (1550 or 1310 nm).
5. Press the
Setup
softkey, then press
SELECT
if necessary to display the Test
Setup menu.
6. Select 1550-nm wavelength, or 1310 nm if option 04.
7. Select the following test setups on the TFS3031:
Fiber Scan: Manual
Test Range: 20 km
Pulse Width: 20 m
Averages: 256
8. Press the
SELECT
button to go to the Format menu.
9. Set the reflectance setting ON.
10. Press the
Exit
softkey.
Summary of Contents for TekRanger 2 TFS3031
Page 5: ......
Page 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 270: ......
Page 271: ......