Performance Check – TekRanger (serial nos. below B052000)
3–12
TFS3031 TekRanger/TekRanger 2 Service Manual
At six-month intervals, measure the reflectance at each wavelength to confirm
the reflectance is 40 dB (
1 dB).
Multimode Dynamic-Range Test Fixture
The multimode dynamic-range test fixture consists of 4.4 to 8.8 kilometers of
62.5/125
µ
m fiber, with connectors and fusion splices as shown below.
62.5/125
µ
m multimode
FC connector
62.5/125
µ
m multimode
ST connector
4.4 to 8.8 km
62.5/125
µ
m
multimode fiber
Denotes splice
Figure 3–5: Multimode Dynamic-Range Test Fixture