Specifications – TekRanger (serial nos. below B052000)
TFS3031 TekRanger/TekRanger 2 Service Manual
1–41
Table 1–44: Horizontal System Dead Zone (Extended Range 1310 nm Singlemode, Option 08)
Characteristic
Performance Requirement
Supplemental Information
Attenuation Dead Zone
Pulsewidth
2000 m
1000 m
500 m
100 m
50 m
20 m
10 m
5 m
1 m
Typical
2300 m
1350 m
600 m
175 m
100 m
40 m
30 m
20 m
45 m
80 m
Attenuation dead zone is defined
as the distance between the initial
and final points, where the
backscatter level is disrupted
beyond ±0.5 dB due to
encountering a single reflective
event.
Based on reflective event with no
greater than -40 dB reflectance.
Based on 9
µ
m, 0.1 NA SM fiber.
Event Dead Zone
Pulsewidth
2000 m
1000 m
500 m
100 m
50 m
20 m
10 m
5 m
1 m
2250 m
1200 m
700 m
115 m
65 m
30 m
20 m
8 m
<5 m
Event dead zone is defined as the
distance between the initial and
final points, where the reflected
pulse amplitude is -1.5 dB down
from its peak value.
Based on reflective event with no
greater than -40 dB reflectance.
Summary of Contents for TekRanger 2 TFS3031
Page 5: ......
Page 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 270: ......
Page 271: ......