Performance Check – TekRanger 2 (serial nos. B052000 and above)
TFS3031 TekRanger/TekRanger 2 Service Manual
4–51
10. Extrapolate the zero-meter backscatter intercept by drawing a straight line on
the backscatter across the incident pulse to the zero-meter point. See figure
4–29. Note where this line intersects the loss axis.
Figure 4–29: Multimode Dynamic-Range Check Zero Intercept, 1-m Pulse Width
11. Press
Zoom Off
to turn off zoom. Press
SELECT
to highlight the cursor
function, then use the arrow keys to move cursor A to 9.6 km.
12. Press and hold the
Cursor A/Cursor B
softkey to join the two cursors.
13. Move cursor B to 9.9 km.
14. Press
Zoom On
. Press
SELECT
to highlight the zoom function, then use
the arrow keys to adjust the zoom so that 400 m of the noise is on the screen.
NOTE
. Half of all acquisitions for this test will not display noise past the end of
the waveform.