Performance Check – TekRanger 2 (serial nos. B052000 and above)
4–44
TFS3031 TekRanger/TekRanger 2 Service Manual
Singlemode Distance and Loss Measurement Accuracy Check
1. Clean the connectors on the TFS3031 and the TFS3031 singlemode loss and
distance accuracy test fixture.
2. Connect the TFS3031 to the loss and distance accuracy test fixture and turn
on power to the instrument.
3. Check that loss at the front-panel connection is minimized (0.75 dB or less).
a. Press the
Setup
softkey to display the Test Setup menu, and set the
range to 1 km and the pulse width to 5 m. Select 1550 wavelength, or
1310 if option 04. Press E
xit
.
b. Press and hold the
START/STOP
button to start a real-time acquisition.
c.
If front-panel loss is greater than 0.75 dB, clean the fiber connector and
the laser output port.
4. Press the
Setup
softkey, then press
SELECT
if necessary to display the Test
Setup menu.
5. Select 1550-nm wavelength, or 1310 nm if option 04.
6. Set the Fiber Scan setup to IntelliTrace.
7. Press the
Exit
softkey.
8. Press the
START/STOP
button to start an acquisition.
9. When the acquisition is complete, verify that the loss and distance reported
on the event table is within tolerance:
Loss must be within
±
0.03 dB (
±
uncertainty of the “calibrated” event)
Distance must be within
±
10 meters (
±
uncertainty of the “calibrated” event)
10. Repeat steps 4 through 9 for 1310 nm if option 06 or 10.
Summary of Contents for TekRanger 2 TFS3031
Page 5: ......
Page 27: ...Module Level Service 1 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 39: ...Battery Recharging and Replacement 1 16 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 129: ...Accessories and Options 1 106 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 133: ...Theory of Operation 2 4 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 265: ...Replaceable Parts 6 14 TFS3031 TekRanger TekRanger 2 Service Manual...
Page 270: ......
Page 271: ......