Acquiring Waveforms
3- 28
TDS5000 Series User Manual
Use the procedure that follows to set up roll mode acquisitions.
Overview
To set roll mode
Control elements and resources
Prerequisites
1.
The horizontal and vertical controls must be set up.
Triggering should also be set up.
Refer to page 3--25 for acquisition setup and
page 3--49 for trigger setup.
Enable roll
mode
2.
Click the Horiz button bar button to open the Horizontal/
Acquisition Setup control window. Select the Acquisition
tab.
3.
Click Roll Mode Auto to enable roll mode.
The resolution, record length, horizontal scale, and
number of active channels on are dependent variables.
Therefore, the fastest horizontal scale setting when Roll
mode starts is 100 ms/div, the record length is 500
(resolution is 2 ms), the acquisition mode is either
Sample or Peak Detect, and 1 or 2 channels must be
on. As the record length or the number of on channels
grows, the time per division required to enter Roll mode
is slower.
Note:
Envelope and Average acquisition modes inhibit
roll mode. Hi Res roll mode starts when the resolution is
4 ms or greater.
Using Roll Mode
Summary of Contents for TDS5054
Page 5: ......
Page 21: ...Preface xvi TDS5000 Series User Manual...
Page 39: ...Installation 1 18 TDS5000 Series User Manual...
Page 165: ...Triggering 3 98 TDS5000 Series User Manual...
Page 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Page 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Page 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Page 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Page 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Page 339: ...Glossary Glossary 12 TDS5000 Series User Manual...