Triggering
TDS5000 Series User Manual
3- 95
Use these procedures to set up the oscilloscope to trigger on a sequence. For
more information, display online help while performing the procedure.
Overview
To trigger on a sequence
Control elements and resources
Prerequisites 1.
The oscilloscope must be installed with a signal
connected to an input channel. Set the acquisition
system to RUN, and set the vertical and horizontal
controls appropriately for the signal to be acquired.
See page 3--21 for acquisition setup
To trigger on
A (main) only
1.
Click Trig on the toolbar, and select the A-->B Seq tab of
the Trigger Setup control window.
2.
Click A Only to turn off sequential triggering.
To trigger on
B after time
1.
To set the time base to run after an A trigger, a trigger
delay, and a B trigger, click Trig on the toolbar and
select the A-->B Seq tab of the Trigger Setup control
window.
2.
Click Trig After Time.
3.
To set the trigger delay, click Trig Delay and use the
multipurpose knob or pop-up keypad to set the time.
4.
To set the B trigger level, click B Trig Level and use the
multipurpose knob or pop-up keypad to set the level.
Triggering on a Sequence
Summary of Contents for TDS5054
Page 5: ......
Page 21: ...Preface xvi TDS5000 Series User Manual...
Page 39: ...Installation 1 18 TDS5000 Series User Manual...
Page 165: ...Triggering 3 98 TDS5000 Series User Manual...
Page 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Page 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Page 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Page 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Page 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Page 339: ...Glossary Glossary 12 TDS5000 Series User Manual...