Triggering
TDS5000 Series User Manual
3- 87
Use this procedure to trigger the oscilloscope when the input signal rises above
an upper threshold level or falls below a lower threshold level.
Overview
To trigger on a window threshold violation
Control elements and resources
Select window
triggering
1.
Push the front-panel ADVANCED button.
2.
On the Trigger Setup control window, select the A Event
tab.
3.
Click Window.
Select the
source
4.
To specify which channel becomes the trigger source,
click Source and select the source from the list.
Set the
thresholds
5.
The upper and lower threshold levels define the voltage
limits of the window. To set the threshold levels, click
Upper Level or Lower Level, and use the multipurpose
knobs or pop-up keypad to set the values.
Triggering on Window
Threshold Violations
Summary of Contents for TDS5054
Page 5: ......
Page 21: ...Preface xvi TDS5000 Series User Manual...
Page 39: ...Installation 1 18 TDS5000 Series User Manual...
Page 165: ...Triggering 3 98 TDS5000 Series User Manual...
Page 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Page 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Page 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Page 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Page 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Page 339: ...Glossary Glossary 12 TDS5000 Series User Manual...