Acquiring Waveforms
3- 36
TDS5000 Series User Manual
H
Interpolation (normal acquisition mode is used instead)
H
Vectors in equivalent-time sampling mode (waveforms are displayed as dots
instead)
H
Saving nonactive waveforms
H
Math waveforms
H
Measurements
Consider the mode you want to use to acquire data:
Automatic Selection.
Fast Acquisitions automatically selects record length and
sample rate to optimize the displayed image by minimizing dead time. Fast
Acquisitions selects sample rates up to 1.25 GS/s and record lengths up to
1,000,000 points, and compresses them to 500 pixels to produce the maximum
display content.
Waveform Capture Rate.
Figures 3--14 and 3--15 illustrate how Fast Acquisitions
mode differs from the normal acquisition mode used by digital storage oscillo-
scopes. Note that normal mode follows a “capture waveform-digitize waveform-
update waveform memory-display waveform” cycle. Normal mode misses events
occurring during the long dead times. Typical waveform capture rates are
50 waveforms per second.
Fast Acquisitions display
Normal DSO display
Figure 3- 14: Normal DSO and Fast Acquisitions displays
Using Fast Acquisitions
Summary of Contents for TDS5054
Page 5: ......
Page 21: ...Preface xvi TDS5000 Series User Manual...
Page 39: ...Installation 1 18 TDS5000 Series User Manual...
Page 165: ...Triggering 3 98 TDS5000 Series User Manual...
Page 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Page 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Page 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Page 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Page 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Page 339: ...Glossary Glossary 12 TDS5000 Series User Manual...