Acquiring Waveforms
3- 44
TDS5000 Series User Manual
Use the procedure that follows to set up FastFrame mode acquisitions.
Overview
To set FastFrame mode
Control elements & resources
Prerequisites
1.
The horizontal and vertical controls must be set up. Set
up the triggering also.
See page page 3--25 for acquisition setup &
page 3--49 for trigger setup.
To set
FastFrame
mode
2.
Click the Horiz button. Select the Acquisition tab from
the Horiz/Acq control window. Click FastFrame Setup to
display the FastFrame Setup control window.
3.
Click FastFrame to toggle FastFrame to On.
Set frame
length
4.
C
lick Rec Length, and set the number of samples per
frame.
Record length is the number of samples in each
acquisition.
To Set FastFrame Mode