Incoming Inspection
TDS5000 Series User Manual
1- 29
Equipment
required
One passive 10X oscilloscope probe (P5050 or equivalent)
One
3.5 inch, 1.44 MB, formatted floppy disk
Prerequisites
None
1.
Initialize the oscilloscope:
Push the front-panel
DEFAULT SETUP
button.
2.
Hook up the signal source:
Connect the oscilloscope probe from the probe
compensation terminal to the CH 1 input as shown in Figure 1--9.
TDS5000 oscilloscope
Connect probe from PROBE
COMP terminal to CH 1 input
Figure 1- 9: Setup for the file system test
3.
Insert the test disk:
Insert the floppy disk in the floppy disk drive at the top
of the front panel.
4.
Set up the oscilloscope:
Push the front-panel
AUTOSET
button.
5.
Set the time base:
Set the HORIZONTAL
SCALE
to
1 ms/div
. The time
base readout is displayed at the bottom of the graticule.
6.
Save the settings:
a.
Select the File menu
Instrument Setup
command. This opens the
Instrument Setup control window.
b.
Click the
Save
button under Save Settings to File. The Save Instrument
Setup As dialog box opens.
c.
From the Save in: drop-down list, select the
3
1
/
2
Floppy (A:)
icon to set
the save destination to the floppy disk.
d.
Note the default file name, and click
Save
to save the setup to the default
file name.
7.
Change the settings again:
Set the HORIZONTAL
SCALE
to
200
s/div
.
File System Test
Summary of Contents for TDS5054
Page 5: ......
Page 21: ...Preface xvi TDS5000 Series User Manual...
Page 39: ...Installation 1 18 TDS5000 Series User Manual...
Page 165: ...Triggering 3 98 TDS5000 Series User Manual...
Page 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Page 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Page 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Page 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Page 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Page 339: ...Glossary Glossary 12 TDS5000 Series User Manual...