specifies that the waveform is transmitted as maximum and minimum point
pairs. Only y values are explicitly transmitted. Absolute coordinates are given
by:
X
n
0
XINcr (n
PT_Off)
Y
nmax
YZEro
YMUlt (y
nmax
YOFf)
Y
nmin
YZEro
YMUlt (y
nmin
YOFf)
specifies a normal waveform where one ASCII or binary data point is
transmitted for each point in the waveform record. Only y values are explicitly
transmitted. Absolute coordinates are given by:
X
n
0
XINcr (n
PT_Off)
Y
n
YZEro
YMUlt (y
n
YOFf)
sets the waveform data point format to enveloped.
WFMPre:PT_Off (No Query Form)
Specifies the trigger point within the waveform record for the reference
waveform specified by the DATa:DESTination command.
Waveform
HORizontal:TRIGger:POsition
= 0 to the recordlength, and is the position of the trigger point relative to
DATa:STARt.
specifies that the trigger point is the first point in the waveform record.
Arguments
Examples
Group
Related Commands
Syntax
Arguments
Examples
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...