![Tektronix TDS 410A Programming Manual Download Page 172](http://html1.mh-extra.com/html/tektronix/tds-410a/tds-410a_programming-manual_1079127172.webp)
Command Descriptions
2–154
TDS Family Oscilloscope Programmer Manual
Table 2–30: Record Length Values (<NR1>)
Configuration
TDS 410A,
TDS 420A,
TDS 460A
TDS 520A,
TDS 524A,
TDS 540A,
TDS 544A
TDS 620A,
TDS 640A,
TDS 644A
TDS 684A
TDS 744A,
TDS 784A
Standard
500,
1000,
2500,
5000,
15000,
30000
500,
1000,
2500,
5000,
15000
500,
1000,
2000
500,
1000,
2500,
5000,
15000
500,
1000,
2500,
5000,
15000,
50000
Option 1M
60000,
120000
50000
75000,
100000,
130000,
250000,
500000
HORIZONTAL:RECORDLENGTH 2500
specifies that 2500 data points will be acquired for each record.
HORIZONTAL:RECORDLENGTH?
might return
15000
as the number of data points per record.
HORizontal:SCAle
Sets the time per division for the main time base and is identical to the HORi-
zontal:MAIn:SCAle command. It is included here for compatibility purposes.
Horizontal
HORizontal:SCAle <NR3>
HORizontal:SCAle?
:
SCAle
?
<Space>
<NR3>
HORizontal
Examples
Group
Syntax
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...