![Tektronix TDS 410A Programming Manual Download Page 188](http://html1.mh-extra.com/html/tektronix/tds-410a/tds-410a_programming-manual_1079127188.webp)
Command Descriptions
2–170
TDS Family Oscilloscope Programmer Manual
MATH1:PROCESSING OFF
ensures that waveform averaging is not in use on math waveform 1.
MATH1:PROCESSING AVERAGE
turns on waveform averaging on math waveform 1.
MEASUrement? (Query Only)
Returns all measurement parameters.
Measurement
MEASUrement?
MEASUrement
?
MEASUREMENT?
might return
:MEASUREMENT:MEAS1:STATE 0;TYPE PERIOD;UNITS
"s";SOURCE1 CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION
FORWARDS;:MEASUREMENT:MEAS2:STATE 0;TYPE PERIOD;UNITS "s";SOURCE1
CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION FORWARDS;
:MEASUREMENT:MEAS3:STATE 0;TYPE PERIOD;UNITS "s";SOURCE1
CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION FORWARDS;
:MEASUREMENT:MEAS4:STATE 0;TYPE PERIOD;UNITS "s";SOURCE1
CH1;SOURCE2 CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION FORWARDS;
:MEASUREMENT:IMMED:TYPE PERIOD;UNITS "s";SOURCE1 CH1;SOURCE2
CH1;DELAY:EDGE1 RISE;EDGE2 RISE;DIRECTION FORWARDS;:MEASUREMENT:
METHOD HISTOGRAM;REFLEVEL:METHOD PERCENT;ABSOLUTE:HIGH 0.0E+0;LOW
0.0E+0;MID 0.0E+0;MID2 0.0E+0;:MEASUREMENT:REFLEVEL:PERCENT:HIGH
90.0E+0;LOW 10.0E+0;MID 50.0E+0;MID2 50.0E+0
MEASUrement:CLEARSNapshot
Removes the measurement snapshot display.
Measurement
MEASUrement:CLEARSNapshot
Examples
Group
Syntax
Examples
Group
Syntax
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...