![Tektronix TDS 410A Programming Manual Download Page 273](http://html1.mh-extra.com/html/tektronix/tds-410a/tds-410a_programming-manual_1079127273.webp)
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–255
TRIGger:MAIn:PULse:RUNT:THReshold:HIGH?
TRIGger
:
:
PULse
MAIn
RUNT
:
:
?
<Space>
<NR3>
HIGH
THReshold
<NR3>
is the threshold, in volts.
TRIGGER:MAIN:PULSE:RUNT:THRESHOLD:HIGH 120E-3
sets the upper limit of the pulse runt trigger to 120 mV.
TRIGger:MAIn:PULse:RUNT:THReshold:LOW
TDS 5XXA, 6XXA, & 7XXA Only
Sets or queries the lower limit for the main pulse runt trigger. This command is
equivalent to setting the threshold in the Pulse Runt Threshold side menu.
Trigger
TRIGger:MAIn:PULse:RUNT:THReshold:LOW <NR3>
TRIGger:MAIn:PULse:RUNT:THReshold:LOW?
THReshold
TRIGger
:
:
PULse
MAIn
RUNT
:
:
?
<Space>
<NR3>
LOW
<NR3>
is the threshold, in volts.
TRIGGER:MAIN:PULSE:RUNT:THRESHOLD:LOW 50E-3
sets the lower limit of the pulse runt trigger to 50 mV.
Arguments
Examples
Group
Syntax
Arguments
Examples
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...