![Tektronix TDS 410A Programming Manual Download Page 271](http://html1.mh-extra.com/html/tektronix/tds-410a/tds-410a_programming-manual_1079127271.webp)
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–253
TRIGger:MAIn:PULse:RUNT:POLarity
TDS 5XXA, 6XXA, & 7XXA Only
Sets or queries the polarity for the main pulse runt trigger. This command is
equivalent to selecting Polarity in the Trigger menu.
Trigger
TRIGger:MAIn:PULse:RUNT:POLarity { EITher | NEGAtive | POSITIVe }
TRIGger:MAIn:PULse:RUNT:POLarity?
NEGAtive
<Space>
?
EITher
POSITIVe
TRIGger
:
:
PULse
MAIn
RUNT
:
POLarity
:
NEGAtive
indicates that the falling edge crosses the high threshold and the rising
edge recrosses the high threshold without either edge ever crossing the low
threshold.
POSITIVe
indicates that the rising edge crosses the low threshold and the falling
edge recrosses the low threshold without either edge ever crossing the high
threshold.
EITher
indicates either
NEGAtive
or
POSITIVe
polarity.
TRIGGER:MAIN:PULSE:RUNT:POLARITY NEGATIVE
specifies that the polarity of the main pulse runt trigger is negative.
TRIGger:MAIn:PULse:RUNT:THReshold? (Query Only)
TDS 5XXA, 6XXA, & 7XXA Only
Returns the upper and lower thresholds for the main pulse runt trigger.
Trigger
TRIGger:MAIn:PULse:RUNT:THReshold?
Group
Syntax
Arguments
Examples
Group
Syntax
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...