Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–201
Table 2–31: Commands that Generate an Operation Complete Message
Operation
Command
Automatic scope adjustment
AUTOSet EXECute
Internal self-calibration
*CAL
Single sequence acquisition
ACQuire:STATE ON or
ACQuire:STATE RUN
(when ACQuire:STOPAfter is set to
SEQuence)
Hardcopy output
HARDCopy STARt
*OPT? (Query Only)
TDS 5XXA, 6XXA, & 7XXA Only
Returns a list of options installed in your digitizing oscilloscope.
Status and Error
*OPT?
*OPT?
OPT?
Might return:
13:Rs232/cent,1M:extended record length,
2F:math pack,05:video trigger,0,CD:color display
.
PASSWord (No Query Form)
Enables the *PUD and NEWpass set commands. Sending PASSWord without
any arguments disables these same commands. Once the password is successful-
ly entered, the *PUD and NEWpass commands are enabled until the digitizing
oscilloscope is powered off, or until the FACtory command, the PASSWord
command with no arguments, or the *RST command is issued.
To change the password, you must first enter the valid password with the
PASSWord command and then change to your new password with the NEWpass
command. Remember that the password is case sensitive.
Miscellaneous
Group
Syntax
Examples
Group
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...