Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–169
!
specifies the number of times to successively average the math
waveform before completing an acquisition.
Successively averages math waveform 2 by 10 times.
might return 10 indicating 10 math 2 waveforms are successively averaged
before a single acquisition occurs.
MATH<x>:PROCessing
TDS 5XXA, 6XXA, & 7XXA Only,
some models require Option 2F
Allows the user to turn on or off averaging for the math waveform specified by
<x>. This is equivalent to selecting No Extended Process or Average in the
Math<x> side menu.
Math averaging allows the digitizing oscilloscope to successively average any
acquisition-related math waveform. This can help reduce noise in a math
waveform.
Vertical
! " # $
!
!
turns off waveform averaging.
turns on waveform averaging.
Arguments
Examples
Group
Syntax
Arguments