Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–247
TRIGger:MAIn:LOGIc:THReshold:CH<x>
TDS 5XXA, 6XXA, & 7XXA Only
Sets or queries the main logic trigger threshold voltage for the channel specified
by <x>. This is equivalent to setting the thresholds in the Logic State Threshold
and Logic Pattern Threshold side menus.
Trigger
TRIGger:MAIn:LOGIc:THReshold:CH<x> <NR3>
TRIGger:MAIn:LOGIc:THReshold:CH<x>?
?
<Space>
<NR3>
TRIGger
:
:
LOGIc
MAIn
:
:
CH
<x>
THReshold
<NR3>
specifies the threshold voltage.
TRIGGER:MAIN:LOGIC:THRESHOLD:CH1 .5
sets the main logic trigger threshold for channel 1 to .5 volts.
TRIGger:MAIn:LOGIc:WHEn
TDS 5XXA, 6XXA, & 7XXA Only
Specifies whether the main logic trigger occurs when the specified state goes true
or false when TRIGger:MAIn:LOGIc:CLAss is set to PATtern. This is equivalent
to selecting Trigger When in the Trigger menu.
Trigger
TRIGger:MAIn:LOGIc:WHEn { FALSe | TRUe }
TRIGger:MAIn:LOGIc:WHEn?
Group
Syntax
Arguments
Examples
Group
Syntax
Summary of Contents for TDS 410A