Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–213
<file path>
<wfm>
SAVe
:
WAVEform
<Space>
REF
<x>
,
<wfm>
is CH<x>, MATH<x>, or REF<x>, and is the waveform that will be
saved.
REF<x>
is the location where the waveform will be stored.
<file path>
(on instruments with the Option 1F File System) is a quoted string
that defines the file name and path. Input the file path using the form
<drive>/<dir>/<filename>
.
<drive>
and one or more
<dir>
s are optional. If
you do not specify them, the digitizing oscilloscope will write the file to the
default directory.
<filename>
stands for a filename of up to 8 characters
followed by a period (“.”) and the proper 3-character extension. Internal format
waveforms use the “
WFM
” extension. The TDS 4XXA and TDS 7XXA can also
use a “
CSV
” extension for spreadsheet format files or a “
DAT
” extension for
MathCad format files. The digitizing oscilloscope will generate an error if you
use any other extension for saving a waveform.
SAVE:WAVEFORM MATH2,REF1
saves the math 2 waveform in reference memory location 2.
SAVE:WAVEFORM MATH1,TEK00000.WFM
saves the math1 waveform to the file
TEK00000.WFM
in the default directory and
on the default drive.
SAVe:WAVEform:FILEFormat
TDS 4XXA & 7XXA Only
Specifies the file format for saved waveforms.
Save and Recall
SAVE:WAVEFORM
SAVe:WAVEform:FILEFormat{ INTERNal | SPREADSheet | MATHCad}
SAVe:WAVEform:FILEFormat?
Arguments
Examples
Group
Related Commands
Syntax
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...