Command Descriptions
2–92
TDS Family Oscilloscope Programmer Manual
DATa:STOP
Sets or queries the last data point that will be transferred when using the
CURVe? query. This allows the transfer of partial waveforms to the controller.
When using the CURVe command, the digitizing oscilloscope will stop reading
data when there is no more data to read or when the specified record length has
been reached so this command will be ignored.
Waveform
CURVe?, DATa SNAp
ranges from 1 to the record length, and is the last data point that will be
transferred. If
is greater than the record length then data will be transferred
up to the record length. If both DATa:STARt and DATa:STOP are greater than
the record length, an execution error will be executed.
If you always want to transfer complete waveforms, just set DATa:STARt to 1
and DATa:STOP to the maximum record length.
When DATa:STOP is less than DATa:STARt, the stop value will equal
DATa:STARt + (DATA:STARt – DATa:STOP). For example, if DATa:STARt =
30 and DATaSTOP = 20, then the range of data points for the waveform transfer
will equal 30 through 40.
specifies that the waveform transfer will stop at data point 15000.
might return
as the last data point that will be transferred.
DATa:TARget
Sets or queries the location for storing waveform data transferred to the
instrument using the CURVe command. This command is equivalent to the
Group
Related Commands
Syntax
Arguments
Examples
Summary of Contents for TDS 410A
Page 4: ......
Page 10: ...Table of Contents vi TDS Family Oscilloscope Programmer Manual ...
Page 56: ...Command Groups 2 38 TDS Family Oscilloscope Programmer Manual ...
Page 346: ...Status and Events 3 20 TDS Family Oscilloscope Programmer Manual ...
Page 352: ...Program Examples 4 6 TDS Family Oscilloscope Programmer Manual ...
Page 358: ...Appendix B Reserved Words B 4 TDS Family Oscilloscope Programmer Manual ...
Page 382: ...Index Index 14 TDS Family Oscilloscope Programmer Manual ...