Appendix C: Calibration and adjustment (RT oscilloscope)
7.
Click
Run
. The green trace should now be circular.
View and correct the
Y-polarization calibration
(RT)
1.
Move the input
fi
ber to get most of the signal on RX channels Y-I and Y-Q.
Tape it down.
2.
Click
Single
on the oscilloscope.
3.
Click
Run
on the OUI.
4.
Observe that the ellipses are displayed on the Constellation plots. Right now
only the Y-constellation has signal. The green trace should line up with the
blue circle in the Y-constellation plot.
5.
If the green trace in Y-Constellation is elliptical:
a.
Click
Stop
.
b.
Enter CorrectY in the separate MATLAB Engine Window.
c.
Copy and paste the resulting pHyb statement before DispCalEllipses,
replacing any other pHyb statement.
6.
Click
Run
. The green trace should now be circular.
OM4000D Series Coherent Lightwave Signal Analyzer
117